The Model 44-116 is a 100 cm² Beta Scintillator that can be used with a several different instruments including survey meters, scalers, ratemeters, and alarm ratemeters

PART NUMBER:47-2696

INDICATED USE: Beta survey
SCINTILLATOR: 0.010" thick plastic scintillation material
WINDOW: Typically 1.2 mg/cm² aluminized mylar
WINDOW AREA:
Active - 125 cm²
Open - 100 cm²
EFFICIENCY (4pi geometry): Typically 15% - Tc-99; 22% - S-90/Y-90; 22% - Cl-36
BACKGROUND:
Typically 30 - 35 cpm/microR/hr
NON-UNIFORMITY: Less than 10%
COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, scalers
TUBE: 1.5"(3.8cm) diameter magnetically shielded photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series "C" (others available)
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 5.5"(13.9cm)H X 4"(10.2cm)W X 12.3"(33cm)L
WEIGHT: 1.5 lb (0.7kg)