The Model 44-6 is a sidewall G-M detector with a rotary beta shield for general purpose survey. |
PART NUMBER:47-1535 |
|
| INDICATED USE: Beta gamma survey DETECTOR: 30 mg/cm² stainless steel wall halogen quenched G-M SENSITIVITY: Typically 1200 cpm/mR/hr (Cs-137 gamma ) BETA CUT OFF: Approximately 200 keV (window open ) ENERGY RESPONSE: Energy dependent DEAD TIME: Typically 95 microseconds COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters, and scalers OPERATING VOLTAGE: 900 volts CONNECTOR: Series "C" (others available ) CONSTRUCTION: 1000 mg/cm² polished stainless steel housing with rotary beta window TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 1.2" (3.0 cm) diameter X 6.5" (16.5 cm)L WEIGHT: 1 lb (0.5kg) |
||
|
|
||
|
Replacement Parts G-M Tube |
||
| Emergency Response | Nuclear Medicine | Radiation Therapy | |
| HOME | |||
| Calibration Services Available | |||
Atlantic Nuclear / tel 781-878-9118 / fax 781 878 3378 / email: anc@att.net 100 Weymouth St. STE E-1 / Rockland, MA 02370 |
|||
|