Model 3-98 Survey Meter

  • Internal Pancake G-M with Beta Shield
  • External Low energy Gamma Scintillator
  • 4 Ranges
  • Total Counting Range from 0 - 200 mR/hr, and 0 - 660,000 cpm

PART NUMBER:48-1135

INDICATED USE: Alpha, beta, low energy gamma survey
DETECTORS: Controlled by selector switch
INTERNAL: Pancake style G-M detector with 1.7 plus or minus 0.3 mg/cm² window
AREA: Active - 15 cm²
Open - 12 cm²
EFFICIENCY(4pi geometry): 5% - C-14; 23% - Sr-90/Y-90; 19% - Tc-99; 33% - P-32; 15% - Pu-239
SENSITIVITY: Typically 3300 cpm/mR/hr (Cs-137 gamma )
ALPHA BETA SHIELD: Retractable 211 mg/cm² phenolic slide
EXTERNAL: Model 44-3 low energy gamma scintillator
EFFICIENCY(4pi geometry): Typically 19% or 675 cpm/microR/hr (I-125)
ENERGY RANGE: Recommended 5 keV - 60 keV
CONNECTOR: Series "C" (others available)
METER DIAL: 0 - 5k cpm, BAT TEST (others available )
MULTIPLIERS: X0.1, X1, X10, X100
LINEARITY: Reading within plus or minus 10% of true value with detector connected
AUDIO: Built in unimorph speaker with ON/OFF switch (greater than 60 dB at 2 feet )
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided )
HIGH VOLTAGE: Adjustable from 200 - 1500 volts
THRESHOLD: 30 mV plus or minus 10 mV
RESPONSE: Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
RESET: Push-button to zero meter
POWER: 2 each "D" cell batteries (housed in sealed compartment that is externally accessible )
BATTERY LIFE: Typically 600 hours with alkaline batteries (battery condition can be checked on meter )
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Cast and drawn aluminum with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 9"(22.9cm)H X 3.5"(8.9cm)W X 8.5"(21.6cm)L
WEIGHT: 5.6 lbs (2.5 kg) including batteries and detector

Common Meter Dials

 202-002
0 - 5k cpm
For Any Detector

 202-608
0-6.6k cpm; 0-2 mR/hr
For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94

 202-664
0-6.6k cpm; 0-20
mSv/h
For Model 44-9; 44-9-18;
44-40; 44-88; 44-89; 44-94

 

 

Accessories

Replacement Parts

Check Source
Scaler Option
Shoulder Strap
Headset
Standard Carrying Case
Air and Water Tight Carrying Case

Cables
Meter Bezel
Meter Movement
Handle
Probe Clip

 

Manual

 
Emergency Response Nuclear Medicine Radiation Therapy

SEARCH

HOME
Calibration Services Available

Atlantic Nuclear / tel 800 878 9118 / fax 781 828 1319 / email: anc@att.net