Model 375-10 Digital Area Monitor

  • LED Digital Display
  • Low and High Alarm Indicators
  • Programmable Alarm Indicator
  • Optional Remote
  • Detector Fail Indicator
  • Battery Backup
  • Data Output/RS-232
  • Internal 2"x2" Nal(Tl) Scintillator


INDICATED USE: Area monitoring
DETECTORS:2"x2" Nal(Tl) Scintillator
DISPLAY:4 digit LED display with 0.8" (2 cm) digits
DISPLAY UNITS:Can be made to display in µR/hr, mR/hr, R/hr,µSv/h, mSv/h, Sv/h, hr, cpm, cps
LINEARITY: Reading within ±10% of true value with detector connected
RESPONSE: Typically 3 seconds from 10% - 90% of final reading
STATUS: (green light)instrument functioning properly
LOW ALARM: Indicated by yellow LOW ALARM light and slow beep (1 per second) audible tone ( can be set at any point from 0.0 - 9999)
HIGH ALARM: Indicated by red HIGH ALARM light and fast beep (4 per second) audible tone (can be set at any point from 0.0 - 9999)
NOTE: Audible alert and alarm indicators can be configured as a single beep if desired
LOW BAT: (yellow) indicates less that 2 hours of battery power remaining
DEAD TIME: Adjustable to compensate for dead time of detector and electronics
OVERLOAD: Senses detector saturation (indicated by display reading "-OL-")
OVERRANGE: Indicates radiation field being measured has exceeded counting range of instrument (indicated by display reading "_____")
DATA OUTPUT: 9 pin connector providing 5 decade logarithmic output, RS-232 output, 10BaseT Ethernet output, signal ground connection, FAIL and ALARM signals (current sink), and direct connection to battery and ground
REMOTE(optional): Model 271, or 272 Remote units
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided)
POWER: 95 - 135 VAC (178 -240 VAC available ), 50-60 Hz single phase (less than 100mA),
6 volt sealed lead-acid rechargeable battery (built in)
BATTERY LIFE: Typically 48 hours in non-alarm condition; 12 hours in alarm condition
BATTERY CHARGER: Battery is continuously trickle charged when instrument is connected to line power and turned on
CONSTRUCTION: Aluminum housing with ivory powder coat paint
TEMPERATURE RANGE: 5° F(-15° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 10.3"(43.2cm)H X 9.7"(24.6cm)W X 3.3"(8.4cm)D
WEIGHT: 9.3 lbs(4.2 kg)

 

OPTIONAL ACCESSORIES

Remote Displays
Relays
Time Date Stamp

 
 

 