The Model 44-25 is a pancake G-M Hand probe for
contamination survey with an alarm ratemeter
PART NUMBER:47-1508
INDICATED USE: Alpha beta gamma survey; Frisking DETECTOR: Pancake type halogen quenched G-M WINDOW: 1.7 plus or minus 0.3 mg/cm² mica WINDOW AREA:
Active - 15 cm²
Open - 12 cm² EFFICIENCY(4pi geometry): Typically 5%-C-14; 22%-Sr-90/Y-90;
19%-Tc-99; 32%-P-32; 15%-Pu-239 SENSITIVITY: Typically 3300 cpm/mR/hr (Cs-137 gamma ) ENERGY RESPONSE: Energy dependent DEAD TIME: Typically 80 microseconds COMPATIBLE INSTRUMENTS: General purpose survey meters, ratemeters,
and scalers OPERATING VOLTAGE: 900 volts CONNECTOR: Series "C" (others available ) CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C) SIZE: 1.8" (4.6 cm)H X 2.7" (6.9 cm)W X 10.7" (27.2 cm)L WEIGHT: 1 lb (0.5kg)