The Model 43-1-1 is a large area dual phosphor alpha/beta scintillator that is designed to be used for simultaneously counting alpha and beta contamination

PART NUMBER:47-2336


INDICATED USE: Alpha-beta survey
SCINTILLATOR: ZnS(Ag) adhered to 0.010" thick plastic scintillation material
WINDOW: Typically 0.8 mg/cm² aluminized mylar (1.2 mg/cm² recommended for outdoor use)
WINDOW AREA:
Active - 83 cm²
Open - 75 cm²
EFFICIENCY (4pi geometry): Typically 30% - Pu-239; 30% - Sr-90/Y-90; 5% - C-14
BACKGROUND:
Alpha - Less than 3 cpm
Beta - Typically 300 cpm or less (10 microR/hr field)
NON-UNIFORMITY: Less than 10%
CROSS TALK:
Alpha to Beta - Less than 10%
Beta to Alpha - Less than 1%
COMPATIBLE INSTRUMENTS: Model 2224, 2225, 2929
TUBE: 1.5"(3.8cm) diameter magnetically shielded photomultiplier
OPERATING VOLTAGE: Typically 500 - 1200 volts
DYNODE STRING RESISTANCE: 100 megohm
CONNECTOR: Series "C" (others available)
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 4.8" (12.2cm) diameter X 9.8" (24.9cm)L
WEIGHT: 2 lb (0.9 kg)

Replacement Parts

Mylar Window
Photomultiplier Tube
Scintillator

 

Emergency Response Nuclear Medicine Radiation Therapy

SEARCH

HOME
Calibration Services Available

Atlantic Nuclear / tel 800 878 9118 / fax 781 828 1319 / email: anc@att.net

1020 Turnpike St. #9 /Canton, MA 02021