Model 3532/3534 Waste Monitors

  • Highly Sensitive Lead Shielded NaI(Tl) Scintillators
  • Red Alarm Strobe
  • Analog Display
  • Collapsible roll around frame

INDICATED USE: Waste monitoring
SENSITIVITY: Will detect a un-shielded 40 microCi Cs-137 source at 10' from the detector and a un-shielded 10 microCi Cs-137 source at 5 feet from the detector
ALARM: Indicated by a red lamp and audible tone (greater than 60 dB at 2 feet full volume )
NOTE: Alarm will automatically reset when contaminated material is removed from the detector area.
METER DIAL: 0 - 20 microR/hr, and 0 - 1.5 kV
LINEARITY: Reading within plus or minus 10% of true value with detector connected
OUTPUT: 102 - 135 VAC (fused 5 Amp) duplex receptacle allowing for connection of alarm strobe, bell, and/or horn
CALIBRATION CONTROLS: Accessible from front of instrument (protective cover provided )
RESPONSE: Dependent on number of counts present, typically 9 seconds from 10% to 90% of final reading
PORTAL OPENING: Adjustable from 36"(91.4cm) - 84"(213.4cm) wide with a height of 77"(195.6cm)
POWER: 95 - 135 VAC (178 - 240 VAC available), 50 - 60 Hz single phase (less than 100 mA)
METER: 2.5" (6.4 cm) arc, 1 mA analog type
CONSTRUCTION: Aluminum housing with beige polyurethane enamel paint
TEMPERATURE RANGE: -4° F(-20° C) to 122° F(50° C)
May be certified for operation from -40° F(-40° C) to 150° F(65° C)
SIZE: 79"(200.7cm)H X 29.6"(75.2cm)D
Minimum Width: 73.2"(185.9cm)
Maximum Width: 121.2"(307.9cm)
SHIPPING SIZE: 96"(243.8cm)H X 48"(121.9cm)W X 36"(91.4cm)D

 

SPECIFICATION

MODEL 3532

MODEL 3534

DETECTORS

2 ea. 0.71" lead shielded 3" diameter X 1" thick NaI scintillation detectors

4 ea. 0.71" lead shielded 3" diameter X 1" thick NaI scintillation detectors

ELECTRONICS

1 ea. Model 3530 electronics

2 ea. Model 3530 electronics

WEIGHT

150 lbs (56.8 kg)

190 lbs (86.4 kg)

SHIPPING WEIGHT

200 lbs (68.2 kg)

340 lbs (155 kg)