| Diagnostic Radiology: Ionization Chambers | ||
|
Flat Chambers 1 cm³, 112 cm³ |
Type 77334, 77335 | |
![]() |
The flat ion chambers 1 cm³ and 112 cm³ are suitable for precise dose
measurements in the useful X-ray beam. The 1 cm³ chamber is calibrated
from 25 kV to 35 kV and from 50 kV to 150 kV with filtering corresponding
to a position behind or in front of a patient.
|
|
|
The 112 cm³ chamber is calibrated from 50 kV to 150 kV with
patient-equivalent filtering. The maximum polarizing voltage for both
chambers is 100 V. The area density of the 1 cm³ chamber is 7.1 mg/cm² and
24 mg/cm² of the 112 cm³ chamber. The chambers can be supplied by a BNT or
TNC or PTW-M connector for operation in connection with a UNIDOS or UNIDOS
E dosemeter. The NOMEX dosemeter requires an M connector. A radioactive
check device is available to correct air density and carry out function
tests.
|
Features:
|
| Flat Chamber 30 cm³ | Type 233612 | |
![]() |
The 30 cm³ flat chamber is especially designed for X-ray dose measurements
in air down to
35 keV radiation energy. The maximum polarizing voltage is 500 V. The area density of the entrance window is 90 mg/cm² . The chamber can be supplied by a BNT or TNC or PTW-M connector. It can be operated in connection with a dosemeter UNIDOS or UNIDOS E. |
|
A radioactive check device including check source and shielding is
available to correct air density and carry out function tests. Using the
appropriate holder, the check source can be positioned and oriented at the
same place on the chamber.
|
Features:
|
| CT Chamber 3.14 cm³ | Type 30009 | |
![]() |
The CT chamber type 30009 is part of the CTDI Set which makes CTDI and
CTDIW determination possible acc. to IEC 61223-2-6
. The chamber can be inserted for measurements into one of five holes of
the CT head or body phantom.
|
Features:
. |
| CT Chamber 4.73 cm³ | Type 77336 | |
![]() |
The long cylindrical 4.73 cm³ CT chamber is designed according to IEC
61223-2-6 for dose length product measurements of CT units. Its special
construction ensures a uniform response over the chamber length. Open
measuring volume, 15 cm long, 0.9 cm outer diameter. Nominal useful range:
30 kV-70 kV.
|
Features:
|